(April 26, 2009) … Part 2 … Immunoassay Testing Video Training Focusses on System Overview of the Tosoh AIA-600 II Automated Immunoassay Analyzer.
Location of Instrument Parts and Reagents: The waste container, and the wash and diluent reservoirs are located on the left side of the instrument. The waste container is a two-liter container. It has a level sensor and two tubings attached to the top of the cap. One is for the waste and one is attached to the vacuum connection. The 1-liter wash and diluent reservoirs also have level sensors to alert the operator when these fluids are low.
The substrate compartment is located on the front left side of the instrument. The instrument does not have a level sensor for the substrate so the amount of substrate in the bottle can be measured using the ruler at the back of the substrate compartment.
The solid waste container is located on the front side of the instrument. The sample compartment is located on the left front of the instrument. The sample chain has 52 routine and 4 stat positions.