(April 26, 2009) … Part 5 … Immunoassay Testing Video Training focusses on Stat Programing and Calibration on the AIA-600 II Automated Immunoassay Analyzer. STAT programming is available when the AIA-600 II is sampling. Pressing the STAT key on the keypad brings up the STAT programming window. The instrument will continue to sample the specimen it is sampling at the time you press STAT. When it is finished the chain will move around until the 4 STAT positions on the chain are in position and when the chain stops moving the STAT programming window will be exposed. Tests are ordered as usual. After the sample ID has been programmed and tests have been requested, press the ASSA START key to begin priority sampling. Once the STAT sample has been processed, the analyzer will resume processing routine samples. Do not remove any specimens or test cups from the sample chain while the analyzer is processing the STAT. It is possible to run only one stat specimen at a time with up to 3 tests ordered on that specimen.